LABORATORIUM MIERNICTWA ELEKTRYCZNEGO
MELLER MARCIN
JASKÓLSKI MARCIN
WOŁYNIAK ROBERT
ĆW NR 2
OBSERWACJA SYGNAŁÓW CYFROWYCH ZA POMOCĄ OSCYLOSKOPU
SPIS PRZYRZĄDÓW :
Oscyloskop radiotechnika OT6620 T1643
Generator Typ POF1 N0 3747 / IV H1
TABELA POMIAROWA:
ΔL=0,05cm
δKu%=1%
δKt%=1%
f |
KU |
LU |
Um |
Um ± ΔUm |
LT |
KT |
T |
f |
f ± Δf |
Hz |
mV/cm |
cm |
mV |
mV |
cm |
s |
s |
Hz |
Hz |
100•103 |
50 |
6,0 |
150 |
150 ± 6 |
5,8 |
2•10-6 |
1,1•10-5 |
86,2•103 |
86,2•103 ± 1,6•103 |
1•103 |
50 |
6,0 |
150 |
150 ± 6 |
5,2 |
2•10-4 |
1,0•10-4 |
961 |
961 ± 18 |
50 |
50 |
6,0 |
150 |
150 ± 6 |
4,3 |
5•10-3 |
2,1•10-2 |
46 |
46 ± 1 |
WZORY I PRZYKŁADY OBLICZEŃ :
(obliczenia wykonano dla f = 100 kHz)
2 Um = KU • LU
Um =
Um = = 150 [ mV ]
ΔKU =
ΔUm = LU ΔKU + KU ΔLU
ΔUm = LU • + KU • ΔLU
ΔUm = 6 • + 50 • 0,05 = 6 [mA]
T = LT • KT
f =
f = [Hz]
f = = 86206,8 = 86,2•103 [Hz]
Δf = +
ΔKT =
Δf = +
Δf =+0,05 = 1605,1=1,6•103[Hz]
WNIOSKI:
Z pomiaru wynika że pomiar napięcia oraz częstotliwości obarczony był dużymi
błędami (szczególnie pomiar częstotliwości), tak duże błędy mogły wynikać z tego że były to pomiary pośrednie.