Cleaning and care of instruments acc.to DIN EN ISO 17664 For dental rotary and surgical instruments

Checking

After cleaning. resp., cleaning/dis infecting. check all instruments for corrosion. damaged surfaces. splintering, distorted shapes (e.g. bent and non-concentiic instruments) as well as contamination and place damaged instruments aside (for the number of times an instrument can be reused. refer to ..Reusability“). Any instruments which are still soiled must be cleaned and disin-fected again.

Care

No maintenance is required.

Do not use instrument oits.

Steel burs must betreated vath anti-corrosion spray.

Wrapping

Arrange the instruments in the sterilising tray before wrapping it in disposable sterilisation wrapping (single or double wrapping) and/or placing it in a sterilisation Container, comptying with the following requirements:

DIN EN 868-2ff/DIN EN ISO/ANSI AAMI ISO 11607

Suitable for steam sterilisation (temperature-resistant up to 137 ®C (279 *F). sufficiently permeable to steam)

Protects the instruments. resp.. sterilisation wrapping adequately against mechanical damage Serviced r eguIar ty as described by the manufacturer (sterilisation Container)

Sterilisation

Sterilisation must only be carried out with the following sterilisation procedures: other sterilisation procedures are not permitted. Steam sterilisation

Gravitation procedurę (after drying the product adequately)

Steam sterilizer as described in DIN EN 13060 or DIN EN 285

Validated according to DIN EN ISO/ANSI AAMI ISO 17665 valid (commissioning and product-specific performance evalu-ation)

Maximum sterilisation temperaturę: 134 *C (273 *F; plus tolerance acc. to DIN EN ISO/ANSI AAMI ISO 17665)

Sterilisation time (exposure time at sterilisation temperaturę) min. 20 min (at 121 C (250 °F)) or. 5 min at 132 °C (270 *F)/134 *C

Proof that the instruni ents are principally suitable for steam sterilisation was prorided by an independent, accredited test labora-tory ustny an H+P Labortechnik (OberschleiRheisn , Gem any) steam steriliser. The procedurę described above was used.

Flash sterilisation is definitely not permitted.

Also do not use hot-air sterilisation. X-ray sterilisation. formaldehyde or ethylene oxide sterilisation. chemiclaves or plasma sterilisation.

Storage

Once sterilised. the instruments must be stored dry and dust-free in their sterilisation wrapping. The instruments must also be protected against sun and heat.

NTUCahla GmbH. Rotary Dental Instruments, kn Camisch 3.07768 Kahla/Germany


2 von 3


Rewsion 2 / gCittig ab Juni 2009



Wyszukiwarka

Podobne podstrony:
gview3 Cleaning and care of instruments acc.to DIN EN ISO 17664 For dental rotary and surgical instr
Performance Schema Event Timing Events are collected by means of instrumentation added to the server
11 3 General rules taking care of the voice •    do not exert your voice for too long
93 Anthropogenic changes of water conditions.. To document long-term trends for flow changes, a curv
A magnetometer is a measuring instrument used to measure the strength and the direction of magnetic
DSC07864 (2) 32 Table 1. Agc correlation of interglacial seas in NW Europę, SE Europę and Baltic are
sprintinst SUPPLEMENTARY INSTRUCTIONS FOR INSTALLATION OF V.W. SPRINT KIT TO:- MICROBUS SPEEDWELL
kryzysy marokańskie (12) >28 CHAPTER U Art. 4: Thcse officcrs and noncommissioned officcrs Jor a
Gaia Mother Gaia, I come on behalf of all humans to apologize for destroying naturę and beg for forg
GSM Networks: Protocols, Terminology, and Implementation BSC takes care of all the central functions
Lec5. Disability terminology. Causes, kinds and grades and social outcomes of disability Introductio

więcej podobnych podstron