LUBUSKIE ZAK£ADY APARATÓW ELEKTRYCZNYCH
LUMEL Spó³ka Akcyjna
ul. Sulechowska 1, 65-950 Zielona Góra
PROGRAMOWALNY
MIERNIK TABLICOWY
NAPIÊCIA I PR¥DU STA£EGO
TYPU
N9
instrukcja obs³ugi
2
SPIS TRECI
1. ZASTOSOWANIE ................................................................. 3
2. ZESTAW MIERNIKA ............................................................. 4
3. RODZAJE WYKONAÑ I KODOWANIE ................................ 4
4. DANE TECHNICZNE ............................................................ 6
5. OPIS KONSTRUKCJI ......................................................... 10
6. INSTALOWANIE ................................................................. 10
7. NASTAWIANIE PARAMETRÓW ........................................ 13
7. 1. Wprowadzanie danych .............................................................
13
7. 2. Nastawianie alarmów ...............................................................
17
7. 3. Obwód wejciowy .....................................................................
18
7. 4. Wyjcie analogowe ...................................................................
19
7. 5. Funkcja przeliczania .................................................................
19
7. 6. Filtrowanie (urednianie) wyników pomiarów ............................
22
7. 7. Ca³kowanie (sumowanie) pomiarów .........................................
22
3
1. ZASTOSOWANIE
Miernik N9 jest przeznaczony jest do pomiaru napiêcia i pr¹du sta³ego
w uk³adach pomiarowych i regulacyjnych. Miernik umo¿liwia
bezporednie pod³¹czenie róde³ sygna³u: napiêcia i pr¹du sta³ego
w zakresie od 60 mV do 1000 V, 200
µ
A do 6 A. Pod³¹czenie boczników
pr¹dowych lub dzielników napiêciowych rozszerza zakresy pomiarowe
do kiloamperów i kilowoltów.
Programowalny miernik N9 realizuje funkcje:
u
pomiaru wielkoci wejciowej i wywietlania na polu cyfrowym,
u
przetwarzania sygna³u mierzonego na analogowy izolowany
galwanicznie sygna³ pr¹du lub napiêcia, do rejestracji lub regulacji,
u
sygnalizacji przekroczenia nastawionych wartoci alarmowych
z mo¿liwoci¹ w³¹czania lub wy³¹czania przekanikami obwodów
zewnêtrznych,
u
pamiêtania maksymalnych i minimalnych wartoci sygna³u mierzonego,
u
komunikacji w sieci cyfrowej przez interfejs RS 485,
u
przeliczania wartoci mierzonych z przekszta³ceniem liniowym lub
nieliniowym np. odczyt napiêcia w zakresie 0...60 mV, a wskazywanie
w jednostkach natê¿enia pr¹du w zakresie 0...1000 A,
u
filtra cyfrowego sygna³u mierzonego (urednianie w krocz¹cym
przedziale czasu),
u
ca³kowania dyskretnego w czasie (sumowanie wyników pomiarów).
Z klawiatury lub poprzez interfejs mo¿na programowaæ zakres
wejciowy, wyjciowy, wprowadzaæ dowoln¹ charakterystykê
przetwarzania oraz programowaæ funkcje
dodatkowe.
4
3. RODZAJE WYKONAÑ I KODOWANIE
Rodzaje wykonañ i kodowanie miernika napiêcia i pr¹du sta³ego
przedstawiono w tablicy 1.
2. ZESTAW MIERNIKA
Do miernika s¹ do³¹czane:
- uchwyty do mocowania w tablicy,
- instrukcja obs³ugi,
- karta gwarancyjna,
- dyskietka z oprogramowaniem dla wykonañ z interfejsem.
W przypadku potrzeby specjalnego wykonania lub bardziej szcze-
gó³owych informacji prosimy o kontakt z Dzia³em Rozwoju firmy
LUMEL S.A., tel. 325 42 51 wew. 53 82.
* dla miernika o zakresie 1000 V maksymalne napiêcie pracy wzglêdem
ziemi ograniczone do 600 V (patrz punkt wymagania bezpie-
czeñstwa).
5
Tablica 1
MIERNIK CYFROWY N9
Zakres miernika
60 mV
01
150 mV
02
200 mV
03
300 mV
04
2 V
05
20 V
06
200 V
07
1000 V*
08
200 mA
09
2 mA
10
20 mA
11
200 mA
12
2 A
13
6 A
14
0 ...20 mA
15
0 ...10 V
16
na zamówienie
99
Wyjcia przekanikowe
bez przekanika
0
1 przekanik
1
2 przekaniki
2
Wyjcie analogowe
bez wyjcia analogowego
0
z wyjciem analogowym pr¹dowym
1
z wyjciem analogowym napiêciowym
2
Interfejs RS 485
bez interfejsu
0
z interfejsem
1
Kolor wywietlacza
czerwony
0
zielony
1
Zasilanie
zasilanie 230 V a.c.
0
zasilanie 127 V a.c.
1
zasilanie 100 V a.c.
2
zasilanie 24 V a.c.
3
Atesty
bez atestów Kontroli Jakoci
0
z atestami Kontroli Jakoci
1
wg uzgodnieñ z odbiorc¹
9
6
4. DANE TECHNICZNE
Pole odczytowe
5 cyfr LED, wysokoæ 14 mm
Zakres wskazañ
-32768 do 32767 programowalny
Zakresy pomiarowe oraz wartoci b³êdów podstawowych przed-
stawiono w tablicy 2.
Wyko-
Zakres miernika
Bocznik
Impedancja
Wskazanie
nanie
wejciowa
na polu
odczytowym
01
-60.00... 60.00 mV
...A/60 mV
≥
20 M
Ω
wg bocznika
0,04
02
-150.00... 150.00 mV
...A/150 mV
≥
20 M
Ω
wg bocznika
0,04
03
-200.00... 200.00 mV
≥
20 M
Ω
± 200.00 mV
0,04
04
-300.0... 300.0 mV
...A/300 mV
≥
20 M
Ω
wg bocznika
0,04
05
-2.0000... 2.0000 V
≥
20 M
Ω
± 2.0000 V
0,04
06
-20.000... 20.000 V
≥
1,8 M
Ω
± 20.000 V
0,04
07
-200.00... 200.00 V
≥
1,8 M
Ω
± 200.00 V
0,04
08 -1000.0... 1000.0 V*
≥
1,8 M
Ω
± 1000.0 V
0,04
09
-200.00... 200.00
µ
A
1 k
Ω
± 200.00
µ
A
0,1
10
-2.0000... 2.0000 mA
100
Ω
± 2.0000 mA
0,1
11
-20.000... 20.000 mA
10
Ω
± 20.000 mA
0,1
12
-200.00... 200.00 mA
1
Ω
± 200.00 mA
0,1
13
-2.0000... 2.0000 A
30 m
Ω
± 2.0000 A
0,1
14
-6.000... 6.000 A
10 m
Ω
± 6.000 A
0,1
15
0... 20.000 mA
10
Ω
± 20.000 mA
0,05
16
0... 10.000 V
≥
1,5 M
Ω
± 10.000 V
0,02
* zakres ograniczony do 600 V d.c.
Tablica 2
B³¹d podstawo-
wy w % górnej
granicy zakresu
pomiarowego
7
Wyjcia analogowe
0...20 mA, 0...10 V ustawiane
w dowolnym przedziale zakresu;
rozdzielczoæ 0,03% górnej granicy
zakresu; dok³adnoæ 0,1%
górnej granicy zakresu.
Wyjcia przekanikowe
2 przekaniki, styki beznapiêciowe
obci¹¿alnoæ 250 V/1 A
Interfejs szeregowy
RS 485, szybkoæ transmisji
9600 bodów
Warunki odniesienia i znamionowe
warunki u¿ytkowania:
- napiêcie zasilania
187...230...244 V
- czêstotliwoæ napiêcia zasilania
48...50...65 Hz
- temperatura otoczenia
0...23...40°C
- wilgotnoæ wzglêdna powietrza
25...75...90%
(niedopuszczalne skroplenia)
- zewnêtrzne pole magnetyczne
0...40...400 A/m
- pozycja pracy
dowolna
- czas nagrzewania
30 min.
- dopuszczalna przeci¹¿alnoæ krótkotrwa³a:
- na zakresach napiêciowych
•
do 2 V
250 V
•
do 1000 V
1500 V
- na zakresach pr¹dowych
10 I
n
/sek
B³¹d dodatkowy w % b³êdu
podstawowego spowodowany:
- wp³ywem temperatury
100%/10°C
- wp³ywem napiêcia zasilania
niezauwa¿alny
Stopieñ ochrony zapewniany
przez obudowê:
wg PN-92/E-08106
- od strony obudowy
IP 40
- od strony zacisków
IP 20
8
Miernik ma fabrycznie wpisane parametry standartowe podane w tablicy 3.
Pobór mocy
< 7 VA
Masa
0,7 kg
Wymiary
48
×
96
×
172 mm
Wymagania bezpieczeñstwa wg PN-IEC 1010-1+A1 (1996)
- izolacja od strony obudowy podwójna
- izolacja miêdzy obwodami podstawowa
Dla napiêæ pracy wzglêdem ziemi do 300 V
- kategoria instalacji III
- stopieñ zanieczyszczenia 2
Dla napiêæ pracy wzglêdem ziemi do 600 V
- kategoria instalacji II
- stopieñ zanieczyszczenia 2
UWAGA!
W celu zastosowania mierników o napiêciu pracy wzglêdem ziemi do
600 V w instalacji kategorii III nale¿y zastosowaæ dodatkowe rodki
bezpieczeñstwa (transformatory separuj¹ce, transoptory itp.)
9
Tablica 3
NAZWA PARAMETRU
NUMER
WARTOÆ FABRYCZNA
PARAMETRU
MAKSYMALNA WARTOÆ
1
DOWOLNA
MINIMALNA WARTOÆ
2
DOWOLNA
ALARM 1
3
20000
HISTEREZA ALARMU 1
4
0
ALARM 2
5
20000
HISTEREZA ALARMU 2
6
0
ZABEZPIECZENIE
7
OFF
STA£A CA£KOWANIA
8
0 sek.
DZIELNIK POMIARÓW
INTEGRAT.
8a
1 000
SETUP
DOLNA WARTOÆ
ZAKRESU WEJ.
1
ZGODNE Z ZAKRESEM
GÓRNA WARTOÆ
ZAKRESU WEJ.
2
ZGODNE Z ZAKRESEM
DOLNA WARTOÆ
ZAKRESU WYJ.
3
00.000 mA
GÓRNA WARTOÆ
ZAKRESU WYJ.
4
20.000 mA
POZYCJA PUNKTU DZIESIÊT.
5
ZGODNE Z ZAKRESEM
FUNKCJE DODATKOWE
6
0
CZAS UREDNIANIA
7
4 sek.
PRZEKSZTA£CENIE
8
LINIOWE y = x
10
Rys.1. Wygl¹d p³yty czo³owej miernika.
W tylnej czêci miernika jest listwa zaciskowa z 9-ma zaciskami
rubowymi oraz 15-to stykowe gniazdo do po³¹czenia roz³¹cznego.
6. INSTALOWANIE
W tablicy nale¿y wyci¹æ otwór o wymiarach jak na rys. 2. Po w³o¿eniu
do otworu, miernik nale¿y umocowaæ za pomoc¹ uchwytów.
Po³¹czenia elektryczne pokazano na rys. 3, a po³¹czenia interfejsu
cyfrowego na rys. 4.
W celu zmniejszenia zak³óceñ na styku zwiernym i rozwiernym
przekaników alarmowych, s¹ do³¹czone uk³ady gasikowe RC,
47 nF + 100
Ω
.
Po do³¹czeniu obci¹¿enia
≤
6 VA mo¿e wyst¹piæ ci¹g³e za³¹czenie
obwodu alarmowego. Wówczas nale¿y przy³¹czyæ, równolegle do
1
N9
2 3
8
AL2
AL1
7
6
5
4
SYGNALIZACJA
ALARMU 1
POLE
WYŒWIETLACZA
SYGNALIZACJA
ALARMU 2
ZNAK " "
–
SYGNALIZACJA
TRYBU ZMIANY
WARTOŒCI
PARAMETRU
NUMER
WYŒWIETLANEJ
WIELKOŒCI
PRZYCISK
ZMNIEJSZANIA
WARTOŒCI
WEJŒCIE
W TRYB ZMIANY
WARTOŒCI
PRZYCISK
ZWIÊKSZANIA
WARTOŒCI
5. OPIS KONSTRUKCJI
Obudowa miernika wykonana z tworzywa sztucznego przewidziana
jest do mocowania w tablicy. Na p³ycie czo³owej rys.1 znajduje siê
pole wywietlaczy, linijka diodowa do wywietlania numeru
parametru, trzy przyciski do nastawiania parametrów oraz diody do
sygnalizacji alarmów.
11
Rys. 2. Wymiary gabarytowo - monta¿owe i wymiary otworu w tablicy.
Rys. 3. Pod³¹czenia zewnêtrzne.
Interfejs
RS485
230 V
50 Hz
U
N
1 2 3 4 5 6 7 8 9
Wejœcie
pr¹dowe
WE I
B
A GND E
WY1
24 V
AL2
AL1
Wyjœcia
przekaŸnikowe
Wyjœcia
ci¹g³e
Wejœcie
napiêciowe
P
min. 105
92
+0,8
45
+0
,6
max. 8
6,2
172
96
48
obci¹¿enia, rezystor o R = 33 k
Ω
, 2 W. Je¿eli nie bêdzie to skuteczne
skontaktowaæ siê z Dzia³em Rozwoju LUMEL S.A.
12
Rys. 4. Sposób pod³¹czenia mierników z interfejsem RS 485
oraz przetwornika 2 przewodowego.
GRUPA MIERNIKÓW 1
MIERNIK 1
MIERNIK 2
MIERNIK n
U
N
A
B
GND
230 V
50 Hz
GRUPA MIERNIKÓW 2
MIERNIK 1+n
MIERNIK 2+n
MIERNIK m+n
U
N
A
B
GND
230 V
50 Hz
EKRAN
RS485
E
E
NASTÊPNA
GRUPA
Interfejs
RS485
230 V
50 Hz
U
N
1 2 3 4 5 6 7 8 9
B
A GND E
AL2
AL1
Wyjœcia
przekaŸnikowe
Wyjœcie
ci¹g³e
Wejœcie
napiêciowe
4... 20 mA
E
13
7. NASTAWIANIE PARAMETRÓW
7. 1. WPROWADZANIE DANYCH
Nastaw dokonuje siê za pomoc¹ przycisków kontroluj¹c przebieg
procesu na polu wywietlaczy. Po w³¹czeniu miernika wykonywany jest
test poprawnoci dzia³ania, a nastêpnie pomiar sygna³u wejciowego.
Pole wskaników parametrów jest wygaszone. Naciniêcie przycisku
powoduje przejcie w tryb wskazywania parametrów pierwszej
grupy, a naciniêcie przycisku
wskazywanie parametrów z drugiej
grupy SETUP, co pokazano na rys. 5 i rys. 6.
PARAMETRY 1... 8
PARAMETRY 1... 8
POMIAR
W£¥CZENIE DO SIECI
1 grupa
2 grupa
SETUP
czas
wyjœcia
60s
czas
wyjœcia
60s
kasowanie alarmu
zero. integratora
Rys. 5. Tryb pracy miernika.
14
AVERAGE
0... 255 s
32 000
ON, OFF
DIV. INTEG.
DZIELNIK POMIARÓW
INTEGRATORA
TRANSFOR
11 POINT
32 000s
INTEGRATE
STA£A CZASOWA
CA£KOWANIA
SECURITY
KOD BEZPIECZEÑSTWA
CZAS UŒREDNIENIA
PUNKTY ODCINKÓW
60 s
60s
60s
60s
60s
60s
60s
±32 000
UWAGA! WartoϾ ujemna
histerezy to dzia³anie
odwrotne alarmu
0 WY£¥CZONE
1 TRANSFORM. W£¥CZ
2... 255 INNE W£¥CZ
HISTER. 2
FUNCTION
0... 255
DEC. POIN.
0... 5
60s
60s
±32 000
ALARM 2
WARTOή ALARMU
POZYCJA PRZECINKA
5
6
7
8
8a
15
SETUP
2
60s
60s
±32 000
HISTER. 1
HISTEREZA ALARMU
60s
60s
±32 000
ALARM 1
WARTOή ALARMU
MAX. OUT.
±32 000
WARTOŒÆ PR¥DU
LUB NAPIÊCIA NA WY
60s
60s
AKTUAL.
MIN. VAL.
MIN. WARTOή
ZMIERZONA
60s
60s
AKTUAL.
MAX. VAL.
MAKS. WARTOή
ZMIERZONA
MIN. OUT.
±32 000
WARTOŒÆ PR¥DU
LUB NAPIÊCIA NA WY
MAX. INP.
±32 000
ZAKRES WEJ.
MAX.
MIN. INP.
±32 000
ZAKRES WEJ.
MIN.
POMIAR WIELKOŒCI
WEJŒCIOWEJ
KASOWANIE
ALARMU; ZERO-
WANIE INTEGRATORA
1
3
4
19999.
Rys. 6. Algorytm obs³ugi miernika.
16
Parametry pierwszej grupy
Tablica 4
Oznaczenie
Numer
Okrelenie parametru
na rys. 6
parametru
MAX. VAL.
1
maksymalna wartoæ podczas pomiarów
MIN. VAL.
2
minimalna wartoæ podczas pomiarów
ALARM 1
3
wartoæ alarmu 1-go
HISTER. 1
4
histereza alarmu 1-go
ALARM 2
5
wartoæ alarmu 2-go
HISTER. 2
6
histereza alarmu 2-go
SECURITY
7
zabezpieczenie przed zmian¹ parametrów
INTEGRATE
8
sta³a czasowa ca³kowania
DIV. INTEG.
8a
dzielnik pomiarów integratora
Parametry drugiej grupy
Tablica 5
Oznaczenie
Numer
Okrelenie parametru
na rys. 6
parametru
MIN. INP
1
dolny zakres wejciowy
MAX. INP
2
górny zakres wejciowy
MIN. OUT
3
dolny zakres wyjciowy
MAX. OUT
4
górny zakres wyjciowy
DEC. POIN.
5
pozycja punktu dziesiêtnego
FUNCTION
6
funkcje dodatkowe
AVERAGE
7
czas uredniania
TRANSFOR.
8
przekszta³cenie wg zadanych punktów
17
Przyciskami i mo¿na nastawiaæ parametry zgodnie
z kolejnoci¹ pokazan¹ na rysunku 6, cyfra podwietlana wskazuje
numer parametru. Aby zmieniæ wartoæ parametru nale¿y nacisn¹æ
przycisk
, przyciskami i nastawiæ po¿¹dan¹ wartoæ.
Tryb zmiany wartoci jest sygnalizowany - patrz rys. 1. Po ustawieniu
wartoci w celu zapamiêtania w pamiêci wewnêtrznej nale¿y powtórnie
nacisn¹æ przycisk
, miernik powróci do trybu wskazywania
parametrów.
7. 2. NASTAWIANIE ALARMÓW
Do sygnalizacji stanów alarmowych zastosowano w mierniku dwa
wyjcia przekanikowe oznaczone jako alarm 1 i alarm 2. Za³¹czenie
lub wy³¹czenie styków przekaników nastêpuje gdy sygna³ mierzony
przekroczy wartoci nastawione dla alarmu 1-go i alarmu 2-go. Histereza
zadzia³ania przekanika jest okrelona wzglêdem nastawionej wartoci
alarmu w dó³ tzn. poni¿ej tej wartoci. Znak + lub - przy histerezie
okrela czy alarm po przekroczeniu nastawionej wartoci bêdzie
w³¹czony czy wy³¹czony.
Dzia³anie przekaników okre laj¹ nastêpuj¹ce parametry:
- wartoæ wielkoci mierzonej przy której 1-szy przekanik zadzia³a,
parametr nr 3 (ALARM1)
- strefa histerezy 1-go przekanika, parametr nr 4 (HISTER 1)
- znak - przy histerezie okrela wy³¹czenie przekanika po przekro-
czeniu nastawionej wartoci alarmu
- znak + przy histerezie okrela w³¹czenie przekanika po przekro-
czeniu nastawionej wartoci
- wartoæ wielkoci mierzonej przy której 2-gi przekanik zadzia³a,
parametr nr 5 (ALARM 2)
- strefa histerezy 2-go przekanika, parametr nr 6 (HISTER 2)
Dzia³anie przekaników ilustruje rys. 7.
18
7. 3. OBWÓD WEJCIOWY
Do mierników N9 mo¿na pod³¹czyæ bezporednio ród³a sygna³u
napiêciowego lub pr¹dowego, oraz w wykonaniu 0...20 mA korzystaj¹c
z wbudowanego zasilacza pod³¹czyæ przetworniki dwuprzewodowe np.
4...20 mA, 1...5 mA. Przyciskami mo¿na zmieniaæ zakres wejciowy
miernika w zale¿noci od potrzeb u¿ytkownika. Przy wspó³pracy
z przetwornikami pomiarowymi np. o zakresie wyjciowym 4...20 mA,
mo¿na ustaliæ zakres wejciowy miernika 0...20 mA na 4 mA do 20 mA,
ale równie¿ na zakres na 6 mA do 15,43 mA. Zakresowi bêdzie
przypisana odpowiednia tablica przetwarzania wartoci mierzonych,
oraz sygna³ wyjciowy.
Rys. 7. Dzia³anie styków przekaników.
ALARM
ALARM
histereza ujemna
histereza dodatnia
przekaŸnik
za³¹czony
przekaŸnik
za³¹czony
przekaŸnik
wy³¹czony
przekaŸnik
wy³¹czony
wartoϾ mierzona
wartoϾ mierzona
0
0
1
1
19
7. 4. WYJCIE ANALOGOWE
Miernik ma wyjcie ci¹g³e z programowanym zakresem w przedziale
0...20 mA lub 0...10 V. Istnieje mo¿liwoæ programowania zakresów
0...2 mA, 1...5 mA, 2...10 mA, 4...20 mA, 1...5 V lub te¿ dowolnych
innych o niestandardowych wartociach np. 11,111...12,345 mA.
Aby ustaliæ ¿¹dany zakres wyjciowy nale¿y podczas
programowania wprowadziæ do komórek o nazwie MIN. OUT. oraz
MAX. OUT. odpowiednie wartoci liczbowe.
Sygna³ analogowy ci¹g³y mo¿e s³u¿yæ w uk³adach automatyki do
rejestracji, regulacji lub przetwarzania. Wyjcie analogowe jest
galwanicznie odizolowane od pozosta³ych czêci uk³adu miernika.
7. 5. FUNKCJA PRZELICZANIA
Funkcja przeliczania mo¿e byæ zastosowana wówczas, gdy do
miernika jest podawany sygna³ standardowy z przetwornika np.
cinienia lub przep³ywu, a na wywietlaczu chcemy odczytywaæ
wartoæ wielkoci mierzonej.
Przetwarzanie odbywa siê w zakresie wczeniej zadeklarowanym
jako zakres wejciowy miernika. Zmiana zakresu wejciowego
wymaga ponownego zdefiniowania wartoci przetwarzanych ze
wzglêdu na minimalizacjê b³êdu przetwarzania.
W³¹czanie funkcji przeliczania nastêpuje poprzez wpisanie cyfry
1 (lub ON) w parametrze FUNCTION w drugiej grupie parametrów
programowania SETUP p. 6.
Aby zaprogramowaæ funkcjê przeliczania nale¿y w parametrze
TRANSFOR. w drugiej grupie parametrów programowania SETUP
p. 8 wprowadziæ 11 - punktow¹ tablicê zale¿noci pomiêdzy war-
tociami zakresu wejciowego sygna³u mierzonego, a wartociami
¿¹danego zakresu wskazañ. Miernik dokona automatycznie podzia³u
zakresu wejciowego zgodnie z wartociami zaprogramowanymi w
komórkach MIN. INP. i MAX. INP. np. dla zaprogramowanego
zakresu wejciowego 0.000....5.000 A zostanie automatycznie
20
ustalony podzia³ 0.000, 0.500, 0.100, 0.150...5.000 A. Nastêpnie
wartociom zakresu wejciowego nale¿y kolejno przyporz¹dkowaæ
wartoci ¿¹danego zakresu wskazañ np. aby dla zaprogramowanego
zakresu 0.000...5.000 A dokonaæ przyporz¹dkowania wartoci
¿¹danych z zakresu 0.000...10.000 A (przek³adnik 10/5 A) nale¿y
wartoci ¿¹danego zakresu podzieliæ na 11 punktów tj. 0.000, 1.000,
2.000, 3.000...10.000 i kolejno przyporz¹dkowaæ wielkociom
zakresu wejciowego. Po zaprogramowaniu funkcji przeliczania
nale¿y ustawiæ pozycjê przecinka dziesiêtnego w SETUP p. 5.
Pozycja przecinka dziesiêtnego nie zmienia rozdzielczoci ani
dok³adnoci wskazañ, a jedynie format wywietlania wartoci.
Przyk³ad:
W mierniku o zakresie wejciowym 0.000...20.000 mA wspó³pracu-
j¹cym z obiektowym przetwornikiem temperatury z wyjciem pr¹-
dowym 4...20 mA zaprogramowaæ zakres wskazañ 0.0...400.0
°
C
zgodnie z zakresem wejciowym przetwornika. W tym celu nale¿y:
zaprogramowaæ doln¹ wartoæ zakresu wejciowego parametr
MIN. INP. SETUP p.1 na wartoæ 04.000 [mA]
zaprogramowaæ górn¹ wartoæ zakresu wejciowego parametr
MAX. INP. SETUP p.2 na wartoæ 20.000 [mA]
zaprogramowaæ pozycjê punktu dziesiêtnego (przecinka) parametr
DEC. POIN. SETUP p.5 na wartoæ 1 (wskazanie 0000.0 zamiast
fabrycznego 00.000)
za³¹czyæ funkcjê przeliczeñ FUNCTION SETUP p.6 ustawiæ
wartoæ 1 (lub ON)
zaprogramowaæ 11 punktow¹ tablicê w parametrze TRANSFOR.
SETUP p.8 wg tablicy 6.
21
Aby wy³¹czyæ zaprogramowan¹ tablicê bez utraty zaprogramowanych w niej
wartoci wystarczy w SETUP p.6 ustawiæ wartoæ 0 (lub OFF).
Uwaga! Miernik z za³¹czon¹ funkcj¹ przeliczania nie pracuje w zakresie
przeci¹¿eniowym. Aby pracowaæ z uwzglêdnieniem tego zakresu nale¿y
zaprogramowaæ tablicê dla pr¹du wejciowego np. 0...6 A.
Tablica 6
WSKAZANIE
CZYNNOÆ
WYWIETLACZA *
0400.0
wcinij ENTER
ustaw wartoæ 0000.0 przyciskami
i ENTER
0560.0
wcinij ENTER
ustaw wartoæ 0040.0 przyciskami
i ENTER
0720.0
wcinij ENTER
ustaw wartoæ 0080.0 przyciskami
i ENTER
0880.0
wcinij ENTER
ustaw wartoæ 0120.0 przyciskami
i ENTER
1040.0
wcinij ENTER
ustaw wartoæ 0160.0 przyciskami
i ENTER
1200.0
wcinij ENTER
ustaw wartoæ 0200.0 przyciskami
i ENTER
1360.0
wcinij ENTER
ustaw wartoæ 0240.0 przyciskami
i ENTER
1520.0
wcinij ENTER
ustaw wartoæ 0280.0 przyciskami
i ENTER
1680.0
wcinij ENTER
ustaw wartoæ 0320.0 przyciskami
i ENTER
1840.0
wcinij ENTER
ustaw wartoæ 0360.0 przyciskami
i ENTER
2000.0
wcinij ENTER
ustaw wartoæ 0400.0 przyciskami
i ENTER
wcinij ENTER KONIEC
* wartoci zakresu wejciowego podzielone automatycznie z przesuniêt¹
pozycj¹ przecinka dziesiêtnego
22
7. 6. FILTROWANIE (UREDNIANIE) WYNIKÓW POMIARÓW
Dla wyeliminowania zak³óceñ lub niestabilnoci w cyklu
pomiarowym miernik zapewnia filtrowanie poprzez urednianie
wyników pomiarów. Aby okreliæ parametry uredniania nale¿y
wprowadziæ podczas programowania miernika liczbê wyra¿on¹
w sekundach dla parametru AVERAGE.
Dzia³anie funkcji jest nastêpuj¹ce:
w zadanym przedziale czasu wyniki pomiarów s¹ sumowane
i dzielone przez liczbê wyników, nastêpnie ze zbioru odrzucany
jest ostatni najstarszy w czasie element (pomiar) a do³¹czany
nowy - najm³odszy. Operacja koñczy siê ponownym obliczeniem
wartoci redniej w przedziale czasu. Cykl jest powtarzany do
momentu zmiany wartoci czasu uredniania. Filtrowanie nosi
cechy obiektu inercyjnego 1-szego rzêdu jakim jest filtr
dolnoprzepustowy zbudowany z elementów RC.
7. 7. CA£KOWANIE (SUMOWANIE) POMIARÓW
Funkcja ca³kowania jest stosowana w przypadku gdy potrzebny
jest pomiar ³adunku elektrycznego np. w galwanizerniach lub te¿
ca³kowania w czasie mierzonych wielkoci np. metrów
szeciennych na godzinê i wskazywania wartoci np. w metrach
szeciennych (m
3
). W celu u¿ycia tej funkcji nale¿y podczas
programowania miernika do komórki o nazwie INTEGRATE
wprowadziæ wartoæ sta³ej czasowej ca³kowania. Wprowadzona
liczba wyra¿ona jest w sekundach tzn. dla liczby 60 sta³a czasowa
ca³kowania wynosi 1 minutê.
23
T - sta³a czasowa ca³kowania
n - liczba wyra¿ona
w sekundach
1
x (t) dt
T
∫
1 ax
i
n
∑
U¿ycie funkcji ca³kowania w po³¹czeniu z ustawieniem wartoci
granicznej (alarm) mo¿e s³u¿yæ do dozowania np. ³adunku
elektrycznego, iloci objêtoci produktu. Aby to zrealizowaæ nale¿y
podczas nastawiania parametrów miernika alarm oraz histereza
wprowadziæ wartoæ przy której maj¹ zadzia³aæ styki przekanika
w³¹czaj¹ce lub wy³¹czaj¹ce obwód zasilania. Wprowadzana liczba
wyra¿ona jest w jednostkach aktualnie mierzonych. Zerowanie wskazañ
nastêpuje przez naciniêcie lewego przycisku klawiatury oznaczony
strza³k¹ w dó³ .
1
DIV. INTEG.
a =
patrz rys. 6